Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/new/cat.php on line 184
General(based on EN 61207/IEC 1207. All data refers to the binary mixture H2 in N2) |
|
Measuring ranges |
4, internally and externally switchable; automatic measurement range switchover also possible |
Largest possible measuring span |
100 vol.% H2(for smallest measuring span, see "Function") |
Measuring ranges with suppressed zero point |
Any zero point within 0 ... 100 vol.% can be implemented, smallest possible measuring span: 5 % H2 |
Operating position |
Front wall, vertical |
Conformity |
CE mark in accordance with EN 61326/A1 and EN 61010/1 |
Design, enclosure |
|
Degree of protection |
IP20 according to EN 60529 |
Weight |
Approx. 10 kg |
Electrical characteristics |
|
EMC (ElectromagneticCompatibility) (All signal lines must be shielded. Measured value deviations of up to 4 % of the smallest measuring range may occur in ranges with strong electromagnetic interference.) |
In accordance with standard requirements of NAMUR NE21 (08/98) |
Electrical safety |
In accordance with EN 61010-1; overvoltage category II |
Power supply (see rating plate) |
100 V -10% ... 120 V +10% AC, 47 ... 63 Hz or 200 V -10% ... 240 V +10% AC, 47 ... 63 Hz |
Power consumption |
Approx. 20 VA |
Fuse values |
100 ... 120 V: 1.0T/250 |
Gas inlet conditions |
|
Sample gas pressure |
800 ... 1 100 hPa (absolute) |
Sample gas flow |
30 ... 90 l/h (0.5 ... 1.5 l/min) |
Sample gas temperature |
Min. 0 to max. 50 °C, but above the dew point |
Temperature of the measuring cell |
Approx. 60 C |
Sample gas humidity |
< 90% relative humidity |
Dynamic response |
|
Warm-up period |
< 30 min (the technical specification will be met after 2 hours) |
Delayed display (T90) |
< 5 s |
Damping (electrical time constant) |
0 ... 100 s, parameterizable |
Dead time (purging time of the gas path in the unit at 1 l/min) |
Approx. 0.5 s |
Measuring response(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
|
Output signal fluctuation |
< ± 0.75% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s ( = 0.25%) |
Zero point drift |
< ± 1%/week of the smallest possible measuring span according to rating plate |
Measured-value drift |
< ± 0.5%/week of the smallest possible span according to rating plate |
Repeatability |
< 1% of the current measuring range |
Detection limit |
1 % of the current measuring range |
Linearity error |
< ± 1% of the current measuring range |
Influencing variable(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
|
Ambient temperature |
< 1%/10 K referred to smallest possible measuring span according to rating plate |
Carrier gases |
Deviation from zero point (for influence of interfering gas see paragraph titled "Interference influences") |
Sample gas flow |
< 0.2% of the smallest possible span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range |
Sample gas pressure |
< 1% of the current measuring range with a pressure change of 100 hPa |
Power supply |
< 0.1% of the current measuring range with rated voltage ± 10% |
Electrical inputs and outputs |
|
Analog output |
0/2/4 ... 20 mA, isolated; |
Relay outputs |
6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated |
Analog inputs |
2, dimensioned for 0/2/4 … 20 mA for external pressure sensor and correction of cross-interference |
Binary inputs |
6, designed for 24 V, isolated, freely parameterizable, e.g. for measurement range switchover |
Serial interface |
RS 485 |
Options |
AUTOCAL function with 8 additional binary inputs and relay outputs each, also with PROFIBUS PA or PROFIBUS DP |
Climatic conditions |
|
Permissible ambient temperature |
-30 … +70 °C during storage and transportation, 5 … 45 °C during operation |
Permissible humidity (dew point must not be undershot) |
< 90% relative humidity as annual average, during storage and transportation |