General information
Areas of application
  • Measurement for boiler control in incineration plants
  • Emission measurements in incineration plants
  • Measurement in the automotive industry (test benches)
  • Warning equipment
  • Process gas concentrations in chemical plants
  • Trace measurements in pure gas processes
  • Environmental protection
  • TLV (Threshold Limit Value) monitoring at the workplace
  • Quality monitoring
  • Ex versions for analyzing flammable and non-flammable gases or vapors for use in hazardous areas
Special versions

Special applications

Besides the standard combinations, special applications concerning material in the gas path, material in the sample cells (e.g. Titan, Hastelloy C22) and measured components are also available on request

TUV version/QAL

TUV-approved versions are available for measurement of CO, NO and SO2 according to 13th and 17th BlmSchV and TA Luft.
Smallest TUV-approved and permitted measuring ranges:

  • 1-component analyzer
    CO: 0 to 50 mg/m3
    NO: 0 to 100 mg/m3
    SO2: 0 to 75 mg/m3
  • 2-component analyzer (series connection)
    CO: 0 to 75 mg/m3
    NO: 0 to 200 mg/m3.
    Furthermore, the TUV-approved versions of the ULTRAMAT 6 comply with the requirements of EN 14956 and QAL 1 in accordance with EN 14181. Conformity of the analyzers with both standards is TUV-certified.
    The analyzer drift can be determined in accordance with EN 14181 (QAL 3) either manually or with a PC using the SIPROM GA maintenance and servicing software. In addition, selected manufacturers of emission evaluation computers offer the possibility for downloading the drift data via the analyzer's serial interface and to automatically record and process it in the evaluation computer.

Flow-type reference compartment

  • The flow through the reference compartment should be adapted to the sample gas flow
  • The gas supply of the reduced flow-type reference compartment should have an upstream pressure of 3 000 to 5 000 hPa (abs.). Then a restrictor will automatically adjust the flow to approximately 8 ml/min






















































Каталог оборудования
Каталог продуктов Siemens Industry Приводная техника Техника автоматизации Системы автоматизации Системы визуализации SIMATIC HMI Системы идентификации Промышленные коммуникации SIMATIC NET Промышленные аппараты управления SIRIUS Промышленные информационные технологии Управление на базе РС Системы управления процессом Датчики Контрольно-измерительные приборы Анализаторы процесса Непрерывные анализаторы газов Непрерывные анализаторы газов, эктрактивные Введение ULTRAMAT 23 ULTRAMAT 6 General information 19" rack unit Field device Documentation Suggestions for spare parts ULTRAMAT/OXYMAT 6 OXYMAT 6 OXYMAT 61 OXYMAT 64 CALOMAT 6 CALOMAT 62 FIDAMAT 6 Непрерывный газовый анализ, in-situ Промышленные газовые хроматографы Analytical Application Sets General information Технологии взвешивания Датчики зрения Блоки питания Продукты для специальных требований Автоматизация зданий Низковольтная коммутационная техника Технология безопасности Системные решения и продукты для отраслей Сервис ... и все, что Вам еще необходимо