Areas of application
- Optimization of small firing systems
- Monitoring of exhaust gas concentration from firing systems with all types of fuel (oil, gas and coal) as well as operational measurements with thermal incineration plants
- Room air monitoring
- Monitoring of air in fruit stores, greenhouses, fermenting cellars and warehouses
- Monitoring of process control functions
- Atmosphere monitoring during heat treatment of steel
- For use in non-potentially-explosive atmospheres
Application areas in biogas plants
- Monitoring of fermenters for generating biogas (input and pure sides)
- Monitoring of gas-driven motors (power generation)
- Monitoring of feeding of biogas into the commercial gas network
Application area of paramagnetic oxygen sensor
- Flue gas analysis
- Inerting plants
- Room air monitoring
- Medical engineering
Further applications
- Environmental protection
- Chemical plants
- Cement industry
Special versions
- Separate gas paths
The ULTRAMAT 23 with 2 IR components without pump is also available with two separate gas paths.. This allows the measurement of two measuring points as used e.g. for the NOx measurement before and after the NOx converter.
The ULTRAMAT 23 gas analyzer can be used in emission measuring systems and for process and safety monitoring. - TUV version/QAL/MCERTS
TUV-approved versions of the ULTRAMAT 23 are available for measurement of CO, NO, SO2 and O2 according to 13th BlmSchV/27th BlmSchV/30th BImSchV (N2O) and TA Luft.
Smallest TUV-approved and permitted measuring ranges:- 1- and 2-component analyzer
CO: 0 to 150 mg/m3
NO: 0 to 100 mg/m3
SO2: 0 to 400 mg/m3 - 3-component analyzer
CO: 0 to 250 mg/m3
NO: 0 to 400 mg/m3
SO2: 0 to 400 mg/m3
- 1- and 2-component analyzer
All larger measuring ranges are also approved.
Furthermore, the TUV-approved versions of the ULTRAMAT 23 comply with the requirements of EN 14956 and QAL 1 according to EN 14181. Conformity of the analyzers with both standards is TUV-certified.
Determination of the analyzer drift according to EN 14181 (QAL 3) can be carried out manually or with a PC using the SIPROM GA maintenance and servicing software. In addition, selected manufacturers of emission evaluation computers offer the possibility for downloading the drift data via the analyzer's serial interface and to automatically record and process it in the evaluation computer.
- Version with reduced response time
The connection between the two condensation traps is equipped with a stopper to lead the complete flow through the measuring cell (otherwise only 1/3 of the flow), i.e. the response time is 2/3 faster. The functions of all other components remain unchanged - Chopper compartment flushing: consumption 100 ml/min (upstream pressure: approx. 3 000 hPa)