19" rack unit

General

Measurement ranges

4, internally and externally switchable; automatic measuring range switchover also possible

Smallest possible span (relating to sample gas pressure 1 000 hPa absolute, 0.5 l/min sample gas flow, and 25 °C ambient temperature)

0 ... 10 vpm O2

Largest possible measuring span

0 ... 100 %

Operating position

Front wall vertical

Conformity

CE mark in accordance with EN 50081-1, EN 50082-2 and RoHS

Design, enclosure

Degree of protection

IP20 according to EN 60529

Weight

Approx. 11 kg

Electrical characteristics

EMC (ElectromagneticCompatibility)

In accordance with standard requirements of NAMUR NE21 (08/98) and EN 61326

Electrical safety

In accordance with EN 61010-1, overvoltage category II

Power supply

100 ... 120 V AC (nominal range of use 90 ... 132 V), 48 ... 63 Hz or

200 ... 240 V AC (nominal range of use 180 ... 264 V), 48 ... 63 Hz

Power consumption

Approx. 37 VA

Fuse values

100 ... 120 V: 1.0T/250

200 ... 240 V: 0.63T/250

Gas inlet conditions

Sample gas flow

  • through the sensor

7.5 l/h

  • Overall consumption

15 ... 30 l/h

Permissible sample gas pressure

  • without internal pressure regulator

2 000 hPa (abs.)

  • with internal pressure regulator

2 000 ... 6 000 hPa (abs.)

Sample gas temperature

Min. 0 ... max. 50 °C, but above the dew point

Sample gas humidity

< 1% relative humidity

Dynamic response

Warm-up period

At room temperature < 30 min (the technical specification will be met after 2 hours)

Damping (electrical time constant)

0 ... 100 s, parameterizable

Dead time (high-pressure version) (purging time of the gas path in the unit at 125 ml/min)

10 ... 30 s

Dead time (low-pressure version without pump)

< 5 s

Dead time (low-pressure version with pump)

< 10 s

Time for device-internal signal processing

< 1 s

Pressure correction range

Pressure sensor internal

800 ... 1 100 hPa (abs.)

Measuring response(referred to sample gas pressure 1 013 hPa absolute, sample gas flow 7.5 l/min, and ambient temperature 25 °C)

Output signal fluctuation

< ± 1% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s

Zero point drift

< ± 1 % of the current span/month

Measured-value drift

< ± 1 % of the current span/month

Repeatability

< 3% of the current measuring span

Detection limit

1 % of current measuring range,

< 0.1 vpm in measuring range

0 ... 10 vpm

Linearity error

< 2% of the current measuring span

Influencing variables(relating to sample gas pressure 1 013 hPa absolute, 7.5 l/min sample gas flow and 25 °C ambient temperature)

Ambient temperature

< 2%/10 K referred to current measuring span

Sample gas pressure only possible if the sample gas can flow out into the ambient air

  • When pressure compensation has been switched off: < 1 % of current span/1 % pressure change
  • When pressure compensation has been switched on: < 0.2 % of current span/1 % pressure change

Residual gases, deviation from zero point

  • Catalytically active sensor (CAZ)

Only gases with non-combustible residual gas components can be introduced

  • Catalytically inactive sensor (CIZ)

Residual gas concentration of 10 vpm H2; CO and CH4 have a lower cross-interference; higher HCs are negligible

Sample gas flow

< 2 % of the smallest possible span with a change in flow of 10 ml/min

Power supply

< 0.1 % of the current measuring range with rated voltage ± 10 %

Electrical inputs and outputs

Analog output

0/2/4 ... 20 mA, 4 … 20 mA (NAMUR), isolated; max. load 750 

Relay outputs

6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated

Analog inputs

2, dimensioned for 0/2/4 … 20 mA for external pressure sensor and correction of influence of residual gas (correction of cross-interference)

Binary inputs

6, designed for 24 V, isolated, freely parameterizable, e.g. for measurement range switchover

Serial interface

RS 485

Options

AUTOCAL function each with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP

Climatic conditions

Permissible ambient temperature

-40 ... +70 °C during storage and transportation, 5 … 45 °C during operation

Permissible humidity

< 90% relative humidity as annual average, during storage and transportation (must not fall below dew point)
























































Каталог оборудования 2012
Каталог продуктов Siemens Industry Приводная техника Техника автоматизации Системы автоматизации Системы визуализации SIMATIC HMI Системы идентификации Промышленные коммуникации SIMATIC NET Промышленные аппараты управления SIRIUS Промышленные информационные технологии Управление на базе РС Системы управления процессом Датчики Контрольно-измерительные приборы Анализаторы процесса Непрерывные анализаторы газов Непрерывные анализаторы газов, эктрактивные Введение ULTRAMAT 23 ULTRAMAT 6 ULTRAMAT/OXYMAT 6 OXYMAT 6 OXYMAT 61 OXYMAT 64 General information 19" rack unit Documentation Suggestions for spare parts CALOMAT 6 CALOMAT 62 FIDAMAT 6 Непрерывный газовый анализ, in-situ Промышленные газовые хроматографы Analytical Application Sets General information Технологии взвешивания Датчики зрения Блоки питания Продукты для специальных требований Автоматизация зданий Низковольтная коммутационная техника Технология безопасности Системные решения и продукты для отраслей Сервис ... и все, что Вам еще необходимо