General information |
|
Power supply |
400 V 3AC +10%/-15%, 50 Hz 400 V 3AC +10%/-15%, 60 Hz Connected load: approx. 5.5 kVA If the delta voltage deviates, a single-phase supply must be provided in addition: 120 V 3AC +10%/-15%, 50 Hz 120 V 3AC +10%/-15%, 60 Hz 230 V 3AC +10%/-15%, 50 Hz 230 V 3AC +10%/-15%, 60 Hz Connected load: approx. 1.5 kVA Other voltages up to 500 V possible on request |
Auxiliary media |
|
|
6 000 ... 8 000 hPa, purified compressed air, free of oil, water and dust |
|
Approx. 4 ... 6 m3/h, depends on purging frequency and duration |
Sample gas connection |
8 mm pipe union, connection for heated or unheated sample gas line; required pump capacity at 700 hPa absolute approx. 2 to 5 l/min |
Liquid-cooled sampling probe |
|
Type |
F6534-B12 |
Material |
Stainless steel, mat. No. 1.4571 |
Length |
1 000/1 500/2 000/2 500/3 000/3 500 mm (corresponds to immersion depth) |
Sampling point |
Dependent on installation:
|
Process temperature |
Up to 1 400 °C |
Coolant |
Synthetic heat transfer liquid |
Coolant flow rate |
Max. 3 200 l/h |
Weight |
Approx. 150 kg |
Electrically heated dust filter |
|
Power supply |
See General information |
Filter |
Sintered metal filter SIKA-R30 (3 m for 98 %) Filters with smaller pore sizes available on request |
Operating temperature |
Approx. 200 °C, isolated contact for low temperature |
Backflushing, two-stage (filter element and filter surface) |
6 000 ... 8 000 hPa, purified compressed air, free of oil, water and dust |
Compressed air connection |
Filter enclosure " Filter pipe " |
Sample gas connection |
|
|
Male thread M24x1.5 |
|
Pipe union DN 8 |
Weight |
Approx. 20 kg |
Compressed air valve manifold |
|
Power supply |
See General information |
Compressed air connection |
6 000 ... 8 000 hPa, purified compressed air, free of oil, water and dust |
Ambient temperature |
Max. 70 °C |
Maximum operating pressure |
16 000 hPa |
Pressure switch |
800 ... 200 hPa absolute, adjustable for detection of low pressure |
Dimensions (WxHxD) in mm |
630 x 380 x 210 |
Weight |
Approx. 40 kg |
Retraction unit with electrical and pneumatic drive |
|
Power supply |
See General information |
Drive |
|
Travel time |
Approx. 90 s |
Dimensions |
Type 1: 3 780 mm for probe length 1 000 ... 1 500 mm Type 2: 5 300 mm for probe length 2 000 ... 3 500 mm |
Weight |
Approx. 420 kg |
Heat exchange unit |
|
Power supply |
See General information |
Heat output |
Max. 65 kW |
Coolant |
Synthetic heat transfer liquid |
Fill quantity |
Approx. 25 l |
Flow |
Max. 3 200 l/h, adjustable |
Operating temperature |
|
|
200 °C |
|
170 °C |
Dimensions (WxHxD) in mm |
1 200 x 1 850 x 1 600 |
Weight |
Approx. 400 kg |
Control and monitoring unit |
|
Power supply |
See General information |
Control voltage |
24 V DC |
Signals |
Floating contacts to the host process control system |
Dimensions (WxHxD) in mm |
760 x 760 x 210 |
Weight |
Approx. 60 kg |