19" rack unit

OXYMAT 61, 19" rack unit

General information

Measuring ranges

4, internally and externally switchable; autoranging is also possible

Smallest possible span(relating to sample gas pressure 1 000 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

2 vol. % or 5 vol. % O2

Largest possible measuring span

100 vol. % O2

Measuring ranges with suppressed zero point

Any zero point within 0 ... 100 vol.% can be implemented, provided that a suitable reference gas is used

Operating position

Front wall, vertical

Conformity

CE mark in accordance with EN 50081-1, EN 50082-2

Design, enclosure

Degree of protection

IP20 according to EN 60529

Weight

Approx. 13 kg

Electrical characteristics

Power supply

100 … 120 V AC (nominal range of use 90 … 132 V), 47 … 63 Hz or
200 … 240 V AC (nominal range of use 180 … 264 V), 47 … 63 Hz

Power consumption

Approx. 37 VA

EMC (ElectromagneticCompatibility)

In accordance with standard requirements of NAMUR NE21 (08/98)

Electrical safety

According to EN 61010-1, overvoltage category III

Fuse values

100 ... 120 V: 1.0 T/250

200 ... 240 V: 0.63 T/250

Gas inlet conditions

Permissible sample gas pressure

  • External reference gas supply

800 … 1 200 hPa absolute

  • With integrated pump

Atmospheric pressure ±50 hPa

Sample gas flow

18 … 60 l/h (0.3 … 1 l/min)

Sample gas temperature

Min. 0 to max. 50 °C, but above the dew point

Sample gas humidity

< 90% relative humidity

Reference gas pressure (high-pressure version)

2 000 ... 4 000 hPa above sample gas pressure, but max. 5 000 hPa absolute (version without reference gas pump)

Reference gas pressure (low-pressure version) with external pump

Min. 100 hPa above sample gas pressure

Dynamic response

Warm-up period

At room temperature < 30 min (the technical specification will be met after 2 hours)

Delayed display (T90)

3.5 s

Damping (electrical time constant)

0 … 100 s, parameterizable

Dead time (purging time of the gas path in the unit at 1 l/min)

Approximately 0.5 ... 2.5 s, depending on version

Time for device-internal signal processing

< 1 s

Pressure correction range

Pressure sensor internal

500 … 2 000 hPa, absolute
(see gas inlet conditions for permissible sample gas pressure)

Measuring response(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

Output signal fluctuation

< ± 0.75% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s (corresponds to ±0.25% at 2 )

Zero point drift

< ± 0.5%/month of the smallest possible span according to rating plate

Measured-value drift

< ±0.5 %/month of the current measuring range

Repeatability

< 1% of the current measuring range

Detection limit

1 % of the current measuring range

Linearity error

< 1% of the current measuring range

Influencing variable(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

Ambient temperature

< 2 %/10 K with span 5%

Sample gas pressure (with air (100 hPa) as internal reference gas supply, correction of the atmospheric pressure fluctuations is only possible if the sample gas can vent to ambient air.)

  • When pressure compensation has been switched off: < 2 % of the current measuring range/1 % pressure change
  • When pressure compensation has been switched on: < 0.2 % of the current measuring range/1 % pressure change

Accompanying gases

Deviation from zero point corresponding to paramagnetic or diamagnetic deviation of accompanying gas (see table)

Sample gas flow at zero point

< 1% of the current measuring range according to rating plate with a change in flow of 0.1 l/min within the permissible flow range

Power supply

< 0.1% of the current measuring range with rated voltage ± 10%

Electrical inputs and outputs

Analog output

0/2/4 … 20 mA, isolated; max. load 750 

Relay outputs

6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, potential-free

Analog inputs

2, dimensioned for 0/2/4 … 20 mA for external pressure sensor and accompanying gas influence correction (correction of cross-interference)

Binary inputs

6, designed for 24 V, isolated, freely parameterizable, e.g. for measuring range switchover

Serial interface

RS 485

Options

AUTOCAL function with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP

Climatic conditions

Permissible ambient temperature

-30 … +70 °C during storage and transportation
5 … 45 °C during operation

Permissible humidity

< 90% relative humidity as annual average, during storage and transportation (must not fall below dew point)
























































Каталог оборудования 2012
Каталог продуктов Siemens Industry Приводная техника Техника автоматизации Системы автоматизации Системы визуализации SIMATIC HMI Системы идентификации Промышленные коммуникации SIMATIC NET Промышленные аппараты управления SIRIUS Промышленные информационные технологии Управление на базе РС Системы управления процессом Датчики Контрольно-измерительные приборы Анализаторы процесса Непрерывные анализаторы газов Непрерывные анализаторы газов, эктрактивные Введение ULTRAMAT 23 ULTRAMAT 6 ULTRAMAT/OXYMAT 6 OXYMAT 6 OXYMAT 61 General information 19" rack unit Documentation Suggestions for spare parts OXYMAT 64 CALOMAT 6 CALOMAT 62 FIDAMAT 6 Непрерывный газовый анализ, in-situ Промышленные газовые хроматографы Analytical Application Sets General information Технологии взвешивания Датчики зрения Блоки питания Продукты для специальных требований Автоматизация зданий Низковольтная коммутационная техника Технология безопасности Системные решения и продукты для отраслей Сервис ... и все, что Вам еще необходимо