C79000-G5200-C143
Внимание! Указаны цены архивного каталога СА01. Для уточнения актуальной цены перейдите по ссылке заказного номера товара или перейдите в каталог СА01 2016
Заказной номер | Вес | PG | Описание продукта | Цена |
C79000-G5200-C143 | 0.42 | 820 | MANUAL ULTRAMAT 6 / OXYMAT 6 | 116.6 |
ULTRAMAT/OXYMAT 6, 19" rack unit |
|
General information |
|
Operating position |
Front wall, vertical |
Conformity |
CE mark in accordance with EN 50081-1, EN 50082-2 |
Design, enclosure |
|
Weight |
Approx. 21 kg |
Degree of protection |
IP20 according to EN 60529 |
Electrical characteristics |
|
EMC (ElectromagneticCompatibility) |
In accordance with standard requirements of NAMUR NE21 (08/98) |
Electrical safety |
According to EN 61010-1, overvoltage category III |
Power supply |
100 ... 120 V AC (nominal range of use 90 ... 132 V), 48 ... 63 Hz or |
Power consumption |
Approx. 70 VA |
Fuse values |
120 ... 120 V: F1/F2 = T 1.6 A |
Electrical inputs and outputs (per channel) |
|
Analog output |
0/2/4 ... 20 mA, isolated; max. load 750 |
Relay outputs |
6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated, non-sparking |
Analog inputs |
2, dimensioned for 0/2/4 … 20 mA for external pressure sensor and correction of influence of accompanying gas (correction of cross-interference) |
Binary inputs |
6, designed for 24 V, isolated, freely parameterizable, e.g. for measuring range switchover |
Serial interface |
RS 485 |
Options |
AUTOCAL function each with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP |
Climatic conditions |
|
Permissible ambient temperature |
-30 ... +70 °C during storage and transportation, 5 … 45 °C during operation |
Permissible humidity |
< 90% relative humidity, during storage and transportation (dew point must not be undershot) |
Technical data, ULTRAMAT channel |
|
Measuring ranges |
4, internally and externally switchable; autoranging is also possible |
Smallest possible measuring range |
Dependent on the application, e.g. |
Largest possible measuring range |
Dependent on the application |
Measuring ranges with suppressed zero point |
Any zero point within 0 ... 100 vol.% can be implemented; smallest possible span 20 % |
Characteristic |
Linearized |
Influence of interfering gases must be considered separately |
|
Gas inlet conditions |
|
Permissible sample gas pressure |
|
|
700 ... 1 500 hPa (absolute) |
|
700 ... 1 300 hPa (absolute) |
Sample gas flow |
18 ... 90 l/h (0.3 ... 1.5 l/min) |
Sample gas temperature |
Min. 0 to max. 50 °C, but above the dew point |
Sample gas humidity |
< 90% (relative humidity), or dependent on measuring task, non-condensing |
Dynamic response |
|
Warm-up period |
At room temperature < 30 min (the technical specification will be met after 2 hours) |
Delayed display (T90-time) |
Dependent on length of analyzer chamber, sample gas line and parameterizable damping |
Damping (electrical time constant) |
0 ... 100 s, parameterizable |
Dead time (purging time of the gas path in the unit at 1 l/min) |
Approx. 0.5 ... 5 s, depending on version |
Time for device-internal signal processing |
< 1 s |
Pressure correction range |
|
Pressure sensor |
|
|
700 ... 1 200 hPa absolute |
|
700 ... 1 500 hPa absolute |
Measuring response(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
|
Output signal fluctuation |
< ± 1% of the smallest possible measuring range according to rating plate |
Zero point drift |
<± 1% of the current measuring range/week |
Measured-value drift |
<± 1% of the current measuring range/week |
Repeatability |
1% of the current measuring range |
Detection limit |
1 % of the smallest possible measuring range |
Linearity error |
< 0.5% of the full-scale value |
Influencing variables(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
|
Ambient temperature |
< 1% of current measuring range/10 K (with constant receiver cell temperature) |
Sample gas pressure |
|
Sample gas flow |
Negligible |
Power supply |
< 0.1% of the current measuring range with rated voltage ± 10% |
Environmental conditions |
Application-specific measuring influences possible if ambient air contains measured component or cross interference-sensitive gases |
Technical data, OXYMAT channel |
|
Measuring ranges |
4, internally and externally switchable; automatic measuring range switchover also possible |
Smallest possible span (relating to sample gas pressure 1 000 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
0.5 vol.%, 2 vol.% or 5 vol.% O2 |
Largest possible measuring range |
100 vol. % O2 |
Measuring ranges with suppressed zero point |
Any zero point within 0 ... 100 vol.% can be implemented, provided that a suitable reference gas is used |
Gas inlet conditions |
|
Permissible sample gas pressure |
|
|
500 ... 3 000 hPa absolute |
|
|
|
500 ... 1 500 hPa absolute |
|
500 ... 1 300 hPa absolute |
Sample gas flow |
18 ... 60 l/h (0.3 ... 1 l/min) |
Sample gas temperature |
0 ... 50 C |
Sample gas humidity |
< 90% RH (relative humidity) |
Reference gas pressure (high-pressure version) |
2 000 ... 4 000 hPa above sample gas pressure, but max. 5 000 hPa |
Reference gas pressure (low-pressure version) |
Min. 100 hPa above sample gas pressure |
Dynamic response |
|
Warm-up period |
At room temperature < 30 min (the technical specification will be met after 2 hours) |
Delayed display (T90 time) |
Min. 1.5 ... 3.5 s, depending on version |
Damping (electrical time constant) |
0 ... 100 s, parameterizable |
Dead time (purging time of the gas path in the unit at 1 l/min) |
Approx. 0.5 ... 2.5 s, depending on version |
Time for device-internal signal processing |
< 1 s |
Pressure correction range |
|
Pressure sensor |
|
|
500 ... 2 000 hPa absolute |
|
500 ... 3 000 hPa absolute |
Measuring response(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
|
Output signal fluctuation |
< 0.75% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s (corresponds to ± 0.25% at 2 ) |
Zero point drift |
< 0.5%/month of the smallest possible measuring span according to rating plate |
Measured-value drift |
0.5%/month of the current measuring range |
Repeatability |
1%/month of the current measuring range |
Detection limit |
1% of the current measuring range |
Linearity error |
1% of the current measuring range |
Influencing variables(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
|
Ambient temperature |
|
Sample gas pressure (with air (100 hPa) as reference gas, correction of the atmospheric pressure fluctuations is only possible if the sample gas can vent to ambient air) |
|
Accompanying gases |
Deviation from zero point corresponding to paramagnetic or diamagnetic deviation of accompanying gas |
Sample gas flow |
< 1% of the smallest possible span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range |
Power supply |
< 0.1 % of the current measuring range with rated voltage ± 10% |