Enclosure | Stainless steel enclosure (1.4301) with the interfaces: - 1 x wall bushing for power supply
- 4 x wall bushing for load cell connection with EMC screw connection
- 4 x wall bushing with blanking plugs
- Ground connection bolt
|
Connection board | Internal connection board - Connection of up to 4 load cells
- Type of analog output
- Type of 24 V direct voltage
|
Integration in automation systems | |
Any automation systems | Via RS 485 (Modbus RTU) |
Communication interfaces | - RS 485 (Modbus RTU)
- 4 digital outputs (24 V DC)
- 4 digital inputs (24 V DC)
- 1 analog output (0/4 ... 20 mA)
|
Commissioning options for the scale | Directly via the color touch panel and the preinstalled "Ready for use" operating software |
Calibration approval | No |
Internal resolution | up to ± 4 million parts |
Number of measurements/second (internal) | 100 Hz |
Filter | - Low-pass filter 0.1 ... 50 Hz
- Mean value filter
|
Weighing functions | |
Weight values | |
Limits | |
Zeroing function | Per command |
Tare function | Per command |
Tare specification | Per command |
Load cells | Strain gauges in 4-wire or 6-wire system |
Load cell excitation | |
Supply voltage (regulated via feedback) | 4.85 V DC |
Permissible load resistance | |
| > 40 ? |
| < 4 100 ? |
With SIWAREX IS Ex interface | |
| > 50 ? |
| < 4 100 ? |
Load cell characteristic | 1 … 4 mV/V |
Permissible range of measuring signal (at greatest set characteristic value) | -21.3 ... +21.3 mV |
Max. distance of load cells | 500 m (229.66 ft) |
Auxiliary power supply | |
Rated voltage | 100 ... 240 V AC |
Line frequency | 50 ... 60 Hz |
Max. power consumption | 0.12 A |
IP degree of protection to DIN EN 60529; IEC 60529 | IP65 |
Climatic requirements Tmin (IND) ... Tmax (IND) (operating temperature) | |
Vertical installation | 0 ... +40 °C (32 ... 104 °F) |
EMC requirements according to | EN 45501 |
Dimensions | 264 x 185 x 97 mm (10.39 x 7.28 x 3.82 in) |
Weight | 4 kg (8.82 lb) |