All electromagnetic flowmeters are based on Faraday's law of induction: UM = B · v · d · k UM = Measured voltage induced in the medium perpendicular to the magnetic field and the flow direction. The voltage is tapped at two point electrodes. B = Magnetic flux density which permeates the flowing medium perpendicular to the flow direction. v = flow velocity of medium d = internal diameter of metering tube k = proportionality factor or sensor constant Function and measuring principle of electromagnetic measurement An electromagnetic flowmeter generally consists of a magnetically non-conducting metering tube with an internal electrically non-conducting surface, magnet coils connected in series and mounted diametrically on the tube, and at least two electrodes which are inserted through the pipe wall and are in contact with the measured medium. The magnet field coils through which the current passes generate a pulsed electromagnetic field with the magnetic flux density B perpendicular to the pipe axis. This magnetic field penetrates the magnetically non-conducting metering tube and the medium flowing through it, which must have a minimum electrical conductivity. According to Faraday's law of induction, a voltage UM is generated in an electrically conducting medium, and is proportional to the flow velocity v of the medium, the magnetic flux density B, and the distance between the electrodes d (internal diameter of pipe). The signal voltage UM is tapped by the electrodes which are in contact with the medium, and passed through the insulating pipe wall. The signal voltage UM which is proportional to the flow velocity is converted by an associated transmitter into appropriate standard signals such as 4 to 20 mA. SITRANS F M diagnosticsThe diagnostic functions are all internal tools in the meter:
SITRANS F M Verificator (MAG 5000 and 6000)The SITRANS F M Verificator is an external tool designed for MAG 5000 and MAG 6000 with MAG 1100, MAG 1100 F, MAG 3100, MAG 3100 P or MAG 5100 W sensors to verify the entire product, the installation and the application. The goal is to improve operation, reduce downtime and maintain measurement accuracy as long as possible. The SITRANS F M Verificator is highly advanced and carries out the complex verification and performance check of the entire flowmeter system, according to unique SIEMENS patented principles. The whole verification test is automated and easy to operate so there is no opportunity for human error or influence. The system is traceable to international standards and tested by WRc (Water Research Council). SITRANS F M Verificator
Verification - StepsVerification of a SITRANS F M flowmeter consists of the following test routines: 1. Transmitter test 1. Transmitter test The transmitter test is the traditional way of on-site testing on the market and checks the complete electronic system from signal input to output. Transmitter test Using the excitation power output, which is generated to drive the magnetic field of the sensor, the verificator simulates flow signal to the transmitter input. By measuring the transmitter outputs the verificator calculates its accuracy against defined values. Test includes:
2. Insulation test Flowmeter insulation test The verification test of the flowmeter insulation is a “cross-talk” test of the entire flowmeter which ensures that the flow signal generated in the sensor is not affected by any external influences. In the “cross-talk” test the verificator generates a high voltage disturbance within the coil circuit and then looks for any “cross-talk” induced in the flow signal circuit. By generating dynamic disturbances close-coupled to the flow signal, the flowmeter is tested for noise immunity to a maximum level:
3. Sensor magnetism test Sensor magnetism test The verification of the sensor magnetism is a “boost” test of the magnetic field coil. The test ensures that the magnetism behaviour is like the first time, by comparing the current sensor magnetism with the “fingerprint” which was determined during initial calibration and stored in the SENSORPROM memory unit. In the “boost” test the verificator changes the magnetic field in certain pattern and with high voltage to get quick stable magnetic condition. This unique test is fulfilled without any interference or compensation of surrounding temperature or interconnecting cabling.
CertificateThe test certificate generated by a PC contains: