Input | |
| Point-to-point connection to a HART field device or Point-to-point connection to a 4 ... 20 mA field device or Up to eight HART field devices with an external voltage supply integrated using multidrop |
Communication | - HART communication with multidrop, as primary or secondary HART master (can be specified)
- 4 ... 20 mA current signal with a point-to-point connection scaling in user-defined measuring range in SITRANS AW210
- Linear
- User-defined scaling with up to 32 points
|
Protocol | HART V7 (compatible with previous HART versions) |
Output | |
Communication | WirelessHART V7 |
Transmission frequency band | 2.4 … 2.4835 GHz (ISM band), 16-channel frequency hopping spread spectrum |
Range (under reference conditions) | Outside up to 235 m (771.00 ft) |
RF signal strength | 10 dBm |
Output signals | |
| - HART Cmd 3
Measured current and up to 4 other dynamic variables (measured values, derived values) or device variables - HART Cmd 9
Up to 8 dynamic variables with status - HART Cmd 48
Additional status information
|
| Scaled or linearized process values |
| - HART Cmd 3
Measured current and up to 4 other dynamic variables (measured values, derived values) or device variables - HART Cmd 9
Up to 8 dynamic variables with status - HART Cmd 48
Additional status information
|
Update time for output signals | You can set the update times separately for the adapter and the connected devices. The possible settings are: - 1, 2, 4, 8, 16, 32 s
- 1, 2, 5, 10, 30, 60 min (times also depend on the gateway)
|
Measuring accuracy | |
Max. measuring error (4 ... 20 mA circuit) | 1 % of measuring range, 40 … 85 °C (104 … 185 °F) |
Rated conditions | |
Location | Outside/inside |
Ambient conditions | |
| -40 ... +85 °C (-40 ... +185 °F) In hazardous areas up to 75 °C (167 °F) |
| -40 ... +85 °C (-40 ... +185 °F) |
Electromagnetic compatibility | To EN 301 489-17 and EN 300 328-1 |
Design | |
Weight | 0.46 kg (1.01 lb) |
Enclosure | |
| |
| Aluminum alloy, RoHS-compliant polyurethane corrosion-resistant coating: |
| Resin |
| ?" NPT female thread |
Degree of protection | IP68 |
Aerial | Potted in enclosure |
Auxiliary power | |
Power supply | Loop power 1 … DC 2.5 V, can be set by user in 0.5 V DC increments |
Loop-powered, operating current | DC 3.2 … 25 mA operating current; overvoltage, surge and reverse polarity protection |
Certificates and approvals | |
Wireless communication approvals | - CE (R&TTE, EMC)
- FCC Part 15.247 for wireless applications in the 2.4 GHz transmission frequency band
- IC
|
Explosion protection | |
Intrinsic safety "i" gases and vapors | II 1G Ex ia IIC T*; IP68 T* = T5 for Ta = -40 ... +85 °C T* = T6 for Ta = -40 ... +75 °C |
Intrinsic safety dust | II 1 D Ex iaD 20 IP68 T95C; Ta = -40 ... +85 °C |
Degree of protection "n" (Zone 2) | II 3 G Ex nA nC IIC T* Gc; IP68 T* = T5 for Ta = -40 ... +85 °C T* = T6 for Ta = -40 ... +75 °C |
Explosion protection acc. to FM for US Intrinsic safe, non-sparking | IS/I,II,III/1/ABCDEFG/ T5 Ta = -40 ... +85 °C, T6 Ta = -40 ... +75 °C NI/I/2/ABCD/ T5 Ta = -40 ... +85 °C, T6 Ta = -40 ... +75 °C S/II,III/2/EFG/ T5 Ta = -40 ... +85 °C, T6 Ta = -40 ... +75 °C I/0/AEx ia/IIC/ T5 Ta = -40 ... +85 °C T6 Ta = -40 ... +75 °C; 20/AEx iaD/T95°C; Ta = -40 ... 85°C I/2/AEx nA nC/IIC/ T5 Ta = -40 ... +85 °C, T6 Ta = -40 ... +75 °C; IP68 |
Explosion protection acc. to FM for CA Intrinsic safe, non-sparking | IS/I,II,III/1/ABCDEFG/ T5 Ta = -40 ... +85 °C T6 Ta = -40 ... +75 °C; NI/I/2/ABCD/ T5 Ta = -40 ... +85 °C T6 Ta = -40 ... +75 °C; S/II,III/2/EFG/ T5 Ta = -40 ... +85 °C T6 Ta = -40 ... +75 °C; I/0/Ex ia/IIC/ T5 Ta = -40 ... +85 °C T6 Ta = -40 ... +75 °C; I/2/Ex nA nC/IIC/ T5 Ta = -40 ... +85 °C T6 Ta = -40 ... +75 °C II/1/EFG Ta = -40 ... +85°C; IP68 |
Flameproof enclosure "d" Gases, vapors | II 2 G Ex d IIC T* Gb; IP68 T* = T5 for Ta = -40 ... +85 °C T* = T6 for Ta = -40 ... +75 °C |
Enclosure protection dust | II 2 D Ex tb IIIC T95°C Ta = -40 ... +85°C; IP68 |
Explosion protection acc. to FM for US Explosion-proof, flame-proof, gas, dust | XP/I/1/ABCD I/1 AEx d IIC T5, T6 Gb DIP/II,III/1/EFG 21/AEx tb IIIC T95°C T5 Ta = -40 ... +85°C, T6 Ta = -40 ... +75°C Type 6P, IP68 |
Explosion protection acc. to FM for CA Explosion-proof, flame-proof, gas, dust | XP/I/1/ABCD I/1 Ex d IIC T5, T6 Gb DIP/II,III/1/EFG T5 Ta = -40 ... +85°C, T6 Ta = -40 ... +75°C |