SITRANS P280 WirelessHART Pressure Transmitter |
Mode of operation | |
Measuring principle | piezo-resistive |
Measured variable | Gauge and absolute pressure |
Gauge pressure input | |
Measuring range | Overload limit/Burst pressure |
0 ... 1.6 bar (0 ... 23 psi) | 4 bar (58 psi) |
0 ... 10 bar (0 ... 145 psi) | 20 bar (290 psi) |
0 ... 50 bar (0 ... 725 psi) | 100 bar (1450 psi) |
0 ... 200 bar (0 ... 2900 psi) | 400 bar (5801 psi) |
0 ... 320 bar (0 ... 4641 psi) | 640 bar (9282 psi) |
Units | mbar, bar, m4H2O, i4H2O, atm, Torr, gcm?, kgcm?, Pa, kPa, MPa, psi, mmHG, mmH2O, ftH2O, inHG, inH2O |
Absolute pressure input | |
Measuring range | Overload limit/Burst pressure |
0 ... 1.6 bar a (0 ... 23 psi a) | 4 bar a (58 psi a) |
0 ... 10 bar a (0 ... 145 psi a) | 20 bar a (290 psi a) |
0 ... 50 bar a (0 ... 725 psi a) | 100 bar a (1450 psia) |
0 ... 200 bar a (0 ... 2900 psia) | 400 bar a (5801 psia) |
0 ... 320 bar a (0 ... 4641 psia) | 640 bar a (9282 psia) |
Units | mbar, bar, m4H2O, i4H2O, atm, Torr, gcm?, kgcm?, Pa, kPa, MPa, psi, mmHG, mmH2O, ftH2O, inHG, inH2O |
Output | |
Output signal | 2.4 GHz Wireless signal with TSMP (Time Synchronized Mesh Protocol) |
Measuring accuracy | According to IEC 60770-1 |
Error in measurement at limit setting including hysteresis and reproducibility | typically 0.17 % of span, max. 0.25 % of sensor span |
Long-term stability | max. ± 0.25 % of sensor/year span |
Effect of ambient temperature | max. ± 0.2 %/10 K of sensor span |
Rated conditions | |
Ambient conditions | |
| -40 ... +80 °C (-40 ... +176 °F) (in ambient temperatures below -20 °C (-4 °F) and above +70 °C (158 °F), readability of the display is limited.) |
| -40 ... +85 °C (-40 ... +185 °F) |
| < 95 % |
Climatic class | 4K4H in accordance with EN 60721-3-4(stationary use at locations not protected against weather) |
Degree of protection | IP65/NEMA 4 |
Permissible medium temperature | -40 ... 85 °C (-40 ... +185 °F) |
Design | |
Enclosure material | low-copper die-cast aluminum, GD-AISi12 (Fe) |
Shock resistance | in accordance with DIN EN 60068-2-29 / 03.95 |
Resistance to vibration | in accordance with EN 60068-2-6/ 12.07 |
Weight | |
| 1.5 kg (3.31 lb) |
| 1.6 kg (3.53 lb) |
Dimensions (W x H x D) | See Dimensional drawing |
Process connection | - G?B male thread to EN 837-1
- ?-14 NPT
|
Sensor break | Is recognized |
Displays and controls | |
Display (with illumination) | |
| 104 x 80 pixels |
| adjustable |
- Number of spaces after comma
| adjustable |
Setting options | - on site with 3 buttons
- with SIMATIC PDM or HART Communicator
|
Auxiliary power | |
Battery | 3.6 V DC |
Communication | |
Radio | WirelessHART V7.1 conforming |
Transmission frequency band | 2.4 GHz (ISM-Band) |
Transmission range under reference conditions | Up to 250 m (line of sight) in outside areas Up to 50 m (greatly dependent on obstacles) in inside areas |
Communication interfaces | - HART communication with HART modem
- WirelessHART
|
Certificates and approvals | |
Wireless communication approvals | R&TTE, FCC |
General product safety | CSAUS/C, CE, UL |
Classification according to pressure equipment directive (PED 2014/68/EU) | Gases: Fluid group 1 Liquids: Fluid group 1; meets requirements as per article 4, paragraph 4 (good engineering practice) |