Non-hazardous area version connection cable - customer interface
Terminal block in the receiver enclosure
Function/voltage
Ethernet cable
1
+
Power supply 19 … 30.2 V, 10 VA1)
2
-
3
Normally closed under power4)
Digital output 0 (relay) 30 V, 0.5 A3)
4
5
Normally closed under power4)
Digital output 1 (relay) 30 V, 0.5 A3)
6
7
+
Digital input 0 0 … 30 V2)
8
-
9
+
Analog output 0 (measurement) 30 V, 24 mA3)
10
-
11
+
Analog output 1 (measurement) 30 V, 24 mA3)
12
-
13
PROFIBUS A line (RxD/TxD_N - data inverted)
Modbus D1 (RxD/TxD_N - data inverted)
RS 485 (PROFIBUS/Modbus)
-7 ... +12 V DC
14
PROFIBUS B line (RxD/TxD_P - data not inverted)
Modbus D0 (RxD/TxD_P - data not inverted)
15
PROFIBUS/Modbus shield
16
Tx+
Ethernet5)
White/orange
17
Tx-
Orange
18
Rx+
White/green
19
Rx-
Green
20
+
Analog input 0 (temperature) 0 … 30 mA2), 120 ?
21
-
22
+
Analog input 1 (pressure) 0 … 30 mA2), 120 ?
23
-
24
Grounding
25
Grounding
Ground
Grounding
Ground
Grounding
Shielding
1) This is the maximum power consumption of the SITRANS SL
2) These are the maximum input values
3) These are the maximum output values
4) Note: "Normal operation" stands for normal operation of the analyzer. The system is connected to the voltage source and is running without problems; no error message generated or displayed. "Normal under power" refers to the status of the relay under the above-named normal operation. The relay contact of the alarm signal is closed.
5) We recommend that the Ethernet connection is not made via the cable to the Ethernet terminals in the detector unit. Instead, the Ethernet connection should be made via the sensor cable connection set which is optionally available for the detector unit.
Examples of digital output and analog output
Example of digital output 0
Example of an analog output 0. Caution: please note that an external isolating power supply may be required!
Sensor cable terminal box on the receiver side (ATEX version)