Densitometer performance available through a density accuracy as follows:
For 316L/1.4404 version better than 0.0007 g/cm3 (0.00025 lb/inch3) with repeatability better than 0.0002 g/cm3 (0.0000072 lb/inch3)
For C22/2.4602 version better than 0.0025 g/cm3 (0.000090 lb/inch3) with repeatability better than 0.0002 g/cm3 (0.0000072 lb/inch3)
One tube without internal welds, reductions or flow splitters offers optimal hygiene, safety and CIP cleanability for food and beverage and pharmaceutical applications.
Larger wall thickness, ensures optimal life-time and corrosion resistance and high-pressure durability
Balanced pipe design with little mechanical energy loss, ensures optimal performance and stability under non-ideal and unstable process conditions (pressure, temperature, density-changes etc.).
4-wire Pt1000 temperature measurement ensures optimum accuracy on mass flow, density and fraction flow
Multi-plug electrical connector & SENSORPROM enable true “plug & play”. Installation and commissioning in less than 10 min.
Intrinsically safe Ex-design ia IIC as standard
Sensor pipe available in high-quality stainless steel AISI 316L/1.4435 or Hastelloy C22/2.4602 offering optimum corrosion resistance.
Rugged and space-saving sensor design in stainless steel matching all applications.
High-pressure program as standard
The sensor calibration factor is also valid for gas measurement.