General information | |
Measuring ranges | 4, internally and externally switchable; manual and autoranging possible |
Smallest possible measuring span | 0 ... 10 vpm |
Largest possible measuring span | 99.999 vpm*) |
Concentration units | vpm, C1, C3, C6 or mgC/m3 |
Autoranging | Hysteresis, selectable |
Measured-value display | Digital concentration display (5 digits with floating point) |
Resolution of digital display | 0.1% of measured value |
Operating position | Front wall, vertical |
Conformity | CE mark in accordance with EN 50081-1, EN 50082-2 |
Oven temperature | Adjustable, 100 ... 200 °C |
Design, enclosure | |
Degree of protection | IP20 according to EN 60529 |
Weight | Approx. 23 kg |
Electrical characteristics | |
Auxiliary power | 100 ... 120 V AC (nominal range of use 90 ... 132 V), 48 ... 63 Hz or 200 ... 240 V AC (nominal range of use 180 ... 264 V), 48 ... 63 Hz |
Power consumption | - Approx. 150 VA during operation,
- approx. 350 VA during warm-up phase
|
EMC interference immunity (electromagnetic compatibility) | In accordance with standard requirements of NAMUR NE21 (08/98) |
Electrical safety | In accordance with EN 61010-1, overvoltage category II |
Fuse values | - 100 ... 120 V: 4.0T/250
- 200 ... 240 V: 2.5 T/250
|
Gas inlet conditions | |
Permissible sample gas pressure | |
| < 2 000 hPa abs. |
| 600 ... 1 100 hPa |
Sample gas flow | 18 ... 60 l/h (0.3 ... 1 l/min) |
Sample gas temperature | 0 ... 200 °C |
Sample gas humidity | < 90% RH (RH: relative humidity) |
Dynamic response | |
Warm-up period | At room temperature, approx. 2 ... 3 h |
Delayed display (T90) | 2 ... 3 s |
Damping (electrical time constant) | 0 ... 100 s, parameterizable |
Dead time (purging time of the gas path in the unit at 1 l/min) | With filter, 2 ... 3 s |
Time for device-internal signal processing | < 1 s |
Measuring response | Based on sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature |
Output signal fluctuation | < 0.75% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s (corresponds to ± 0.25% at 2 ?) |
Zero point drift | < 0.5%/month of the smallest possible measuring span according to rating plate |
Measured-value drift | < 1%/week of the current measuring range |
Repeatability | < 1% of the current measuring range |
Detection limit | 0.1 vpm (version for ultra-pure gas measurement: 50 ppb) |
Linearity error | < 1% of the current measuring range |
Influencing variables | Based on sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature |
Ambient temperature | < 1%/10 K referred to smallest possible span according to rating plate |
Atmospheric pressure | < 1%/50 hPa |
Sample gas pressure | < 2% of the current measuring range range/1% pressure change (within 600 ... 1 100 hPa) |
Auxiliary power | < 1% of the current measuring range with rated voltage ± 10% |
Position influence | < 1% with < 15° inclination |
Electrical inputs and outputs | |
Analog output | 0/2/4 ... 20 mA, isolated; max. load 750 ? |
Relay outputs | 6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, potential-free |
Analog inputs | 2, dimensioned for 0/2/4 to 20 mA for external pressure sensor and correction of influence of accompanying gas (correction of cross-interference) |
Digital inputs | 6, designed for 24 V, floating, freely parameterizable, e.g. for measuring range switchover |
Serial interface | RS 485 |
Options | AUTOCAL function each with 8 additional digital inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP |
Climatic conditions | |
Permissible ambient temperature | 5 ... 45 °C in operation, -30 ... +70 °C during storage and transportation |
Permissible humidity | < 90% RH (RH: relative humidity) as annual average, during storage and transportation (must not fall below dew point) |