General information19" rack unit- With 4 HU for installation
- in hinged frame
- In cabinets with or without telescope rails
- Front plate can be swung down for servicing purposes (laptop connection)
- Internal gas paths: hose made of FKM (Viton) or pipe made of titanium or stainless steel (mat. no. 1.4571)
- Gas connections for sample gas inlet and outlet and for reference gas: fittings, pipe diameter of 6 mm or ?"
- Flow indicator for sample gas on front plate (option)
- Pressure switch in sample gas path for flow monitoring (option)
Field device- Two-door enclosure with gas-tight separation of analyzer and electronics sections
- Each half of the enclosure can be purged separately
- Analyzer unit and piping can be heated up to 130 °C (option)
- Gas path and stubs made of stainless steel (mat. no. 1.4571) or titanium, Hastelloy C22
- Purging gas connections: pipe diameter 10 mm or 3/8"
- Gas connections for sample gas inlet and outlet and for reference gas: clamping ring connection for a pipe diameter of 6 mm or ?"
Display and control panel- Large LCD panel for simultaneous display of:
- Measured value (digital and analog displays)
- Status bar
- Measuring ranges
- Contrast of LCD panel adjustable using menu
- Permanent LED backlighting
- Washable membrane keyboard with five softkeys
- Menu-driven operation for parameterization, test functions, adjustment
- User help in plain text
- Graphic display of concentration trend; programmable time intervals
- Bilingual operating software German/English, English/Spanish, French/English, Spanish/English, Italian/English
Input and outputs- One analog output per measured component (from 0, 2, 4 to 20 mA; NAMUR configurable)
- Two analog inputs configurable (e.g. correction of cross-interference, external pressure sensor)
- Six digital inputs freely configurable (e.g. for measurement range switchover, processing of external signals from sample preparation)
- Six relay outputs freely configurable (failure, maintenance request, maintenance switch, threshold alarm, external magnetic valves)
- Expansion: Eight additional digital inputs and eight additional relay outputs each e.g. for autocalibration with up to four calibration gases
CommunicationRS 485 present in the basic unit (connection at the rear; for the rack unit also behind the front plate). Options - AK interface for the automotive industry with extended functions
- RS 485/RS 232 converter
- RS 485/Ethernet converter
- RS 485/USB converter
- Connection to networks via PROFIBUS DP/PA interface
- SIPROM GA software as the service and maintenance tool
 OXYMAT 6, membrane keyboard and graphic display Designs – Parts touched by sample gas, standardGas path | 19" rack unit | Field device | Field device Ex |
|---|
With hoses | Bushing | Stainless steel, mat. no. 1.4571 | - | - | Hose | FKM (e.g. Viton) | Sample chamber | Stainless steel, mat. no. 1.4571 or Tantalum | Fittings for sample chamber | Stainless steel, mat. no. 1.4571 | Restrictor | PTFE (e.g. Teflon) | O-rings | FKM (e.g. Viton) | With pipes | Bushing Pipe Sample chamber Restrictor O-rings | Titanium Titanium Stainless steel, mat. no. 1.4571 or Tantalum Titanium FKM (Viton) or FFKM (Kalrez) | With pipes | Bushing Pipe Sample chamber Restrictor O-rings | Stainless steel, mat. no. 1.4571 Stainless steel, mat. no. 1.4571 Stainless steel, mat. no. 1.4571 or tantalum Stainless steel, mat. no. 1.4571 FKM (Viton) or FFKM (Kalrez) | With pipes | Bushing Pipe Sample chamber Restrictor O-rings | | Hastelloy C 22 Hastelloy C 22 Stainless steel, mat. no. 1.4571 or tantalum Hastelloy C 22 FKM (e.g. Viton) or FFKM (e.g. Kalrez) |
OptionsFlow indicator | Measurement pipe Variable area Suspension boundary Angle pieces | Duran glass Duran glass, black PTFE (Teflon) FKM (Viton) | - | - | Pressure switch | Membrane Enclosure | FKM (Viton) PA 6.3 T | - | - |
Gas path (19" rack unit)Legend for the gas path figures | | |
|---|
1 | Sample gas inlet | 8 | Pressure switch in sample gas path (option) | 2 | Sample gas outlet | 9 | Purging gas | 3 | Not used | 10 | Pressure switch in reference gas path (option) | 4 | Reference gas inlet | 11 | Pressure sensor | 5 | Restrictor in reference gas inlet | 12 | Filter | 6 | O2 physical system | 13 | Flow indicator in sample gas path (option) | 7 | Restrictor in sample gas path | 14 | Outlet restrictor |
 Gas path, reference gas connection 1 100 hPa, absolute  Gas path, reference gas connection 3 000 to 5 000 hPa, absolute Gas path (field device)Legend for the gas path figures | | |
|---|
1 | Not used | 8 | Purging gas inlet (analyzer side) | 2 | Sample gas inlet | 9 | Pressure sensor | 3 | Reference gas inlet | 10 | O2 physical system | 4 | Sample gas outlet | 11 | Restrictor in sample gas path | 5 | Purging gas inlet (electronics side) | 12 | Pressure sensor in reference gas path (option) | 6 | Purging gas outlet (electronics side) | 13 | Restrictor | 7 | Purging gas outlet (analyzer side) | 14 | Outlet restrictor |
 Gas path, reference gas connection 1 100 hPa, absolute  Gas path, reference gas connection 3 000 to 5 000 hPa, absolute
|
|