Measuring ranges | 4, internally and externally switchable; automatic measuring range switchover also possible |
Smallest possible span (relating to sample gas pressure 1 000 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) | 0.5 vol.%, 2 vol.% or 5 vol.% O2 |
Largest possible measuring range | 100 vol.% O2 |
Measuring ranges with suppressed zero point | Any zero point within 0 ... 100 vol.% can be implemented, provided that a suitable reference gas is used |
Gas inlet conditions | |
Permissible sample gas pressure | |
| 500 ... 3 000 hPa absolute |
| |
| 500 ... 1 500 hPa absolute |
| 500 ... 1 300 hPa absolute |
Sample gas flow | 18 ... 60 l/h (0.3 ... 1 l/min) |
Sample gas temperature | 0 ... 50 ?C |
Sample gas humidity | < 90% RH (relative humidity) |
Reference gas pressure (high-pressure version) | 2 000 ... 4 000 hPa above sample gas pressure, but max. 5 000 hPa |
Reference gas pressure (low-pressure version) | Min. 100 hPa above sample gas pressure |
Dynamic response | |
Warm-up period | At room temperature < 30 min (the technical specification will be met after 2 hours) |
Delayed display (T90-time) | Min. 1.5 ... 3.5 s, depending on version |
Damping (electrical time constant) | 0 ... 100 s, configurable |
Dead time (purging time of the gas path in the unit at 1 l/min) | Approx. 0.5 ... 2.5 s, depending on version |
Time for device-internal signal processing | < 1 s |
Pressure correction range | |
Pressure sensor | |
| 500 ... 2 000 hPa absolute |
| 500 ... 3 000 hPa absolute |
Measuring response | Based on sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature |
Output signal fluctuation | < 0.75% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s (corresponds to ± 0.25% at 2?) |
Zero point drift | < 0.5%/month of the smallest possible measuring span according to rating plate |
Measured-value drift | ? 0.5%/month of the current measuring range |
Repeatability | ? 1%/month of the current measuring range |
Detection limit | 1% of the current measuring range |
Linearity error | 1% of the current measuring range |
Influencing variables | Based on sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature |
Ambient temperature | - < 0.5%/10 K referred to smallest possible span according to rating plate
- With measuring span 0.5%: 1%/10 K
|
Sample gas pressure (with air (100 hPa) as reference gas, correction of the atmospheric pressure fluctuations is only possible if the sample gas can vent to ambient air) | - With disabled pressure compensation: < 2% of the current measuring range /1 % change in atmospheric pressure
- With disabled pressure compensation: < 0.2% of the current measuring range /1 % change in atmospheric pressure
|
Accompanying gases | Deviation from zero point corresponding to paramagnetic or diamagnetic deviation of accompanying gas |
Sample gas flow | < 1% of the smallest possible span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range |
Auxiliary power | < 0.1% of the current measuring range with rated voltage ± 10% |