Installation guidelines MC2 DN 100 and DN 150

Installation of sensor

Rigid mounting brackets must be used when installing the sensor. The brackets must be installed as close to the sensor as possible, attached to the piping outside the process connections.

The optimal installation orientation is a vertical installation with an upward flow as shown in the following figure. This has the advantage that any solids contained in the fluid will settle downward and gas bubbles will move upward out of the meter tube when the flow rate is zero. Additionally, it is easy to drain the meter tube. Deposits can thereby be avoided.

Vertical orientation:

Vertical installation self-draining (upward flow)

Horizontal orientation, self-draining

Avoid vibrations

Avoid cross talk

Installation in a drop line

Mount with reduction (A) or orifice (B) to prevent partially draining (min. back pressure: 0.2 bar).

Installation in a drop line