Measuring range | 1 ... 3 300 pF |
Span | Min. 3.3 pF |
Output | |
Solid-state switch | |
| Galvanically isolated |
| Bipolar |
| |
| 82 mA |
| < 1 V, typical at 50 mA |
- Time delay (pre or post switching)
| 1 ... 60 s |
Loop current | 3.6 ... 22 mA/22 ... 3.6 mA (2-wire current loop) |
Accuracy (transmitter) | |
Temperature stability
| 0.15 pF (0 pF) or < 0.25 % (typically < 0.1 %) of actual measured value, whichever is greater over the full temperature range |
Non-linearity and repeatability | < 0.1 % of range and actual measured value respectively |
Accuracy | Deviation < 0.1 % of measured value |
Rated operating conditions1) | |
Installation conditions | |
Location | Indoor/outdoor |
Ambient conditions | |
Ambient temperature (transmitter) | -40 ... +85 °C (-40 ... +185 °F)2) |
Installation category | II |
Pollution degree | 4 |
Medium conditions | |
Relative dielectric constant ?r | Min. 1.5 |
Minimum difference in dielectric constant for interface | 5 |
Process temperature | Temperature rating of process seal is pressure dependent. See Pressure/Temperature curves on page 5/292. |
| -50 ... +200 °C (-58 ... +392 °F) |
| -200 ... +200 °C (-328 ... +392 °F) |
Process pressure | Pressure rating of process seal is temperature dependent. See Pressure/Temperature curves on page 5/292. |
Standard (PFA) | -1 ... 150 bar g (2 175 psi g) |
Design | |
Material | |
| |
| 316L stainless steel |
| PFA |
| 316 stainless steel/ 316 stainless steel PFA |
Probe diameter | |
| 16 mm (0.63 inch) or 24 mm (0.95 inch) |
| 9 mm (0.35 inch) with PFA jacket, 6 mm (0.24 inch) without PFA jacket |
Active shield length | |
| 50 mm (1.97 inch), customer selectable (order number Y02) |
Probe length | |
| Max. 3.5 m (138 inch) with 16 mm rod, PFA Max. 5.5 m (216 inch) with 24 mm rod, PFA |
| Max. 35 m (1 378 inch) |
Process connection of probe | |
| NPT [(Taper), ANSI/ASME B1.20.1] R [(BSPT), EN 10226/PT (JIS-T), JIS B 0203] G [(BSPP), EN ISO 228-1/PF (JIS-P), JIS B 0202] |
| ASME, EN 1092-1 |
Enclosure | |
Material | Aluminum, epoxy-coated |
Cable inlet | 2 x ? inch NPT (2 x M20 x 1.5, IP68 adapter, optional) |
Degree of protection | Type 4X/NEMA4X/IP65, IP68 |
Power supply | 12 ... 33 V DC |
User Interface | |
Display | Local LCD, 4 digit, each 0 ... 9 and limited alpha characters |
Rotary function switch | For selecting programmable menu items |
Push buttons | Red +, blue -, used in conjunction with rotary switch for programming |
Features | |
Measurement current signaling | According to NAMUR NE 43, signal 3.8 ... 20.5 mA, fault ? 3.6 or ? 21 mA (22 mA) |
Safety | - Inputs/outputs fully galvanically isolated
- Polarity-insensitive current loop
- Fully potted
- Integrated safety barrier
|
- Diagnostics with fault alarm when:
| Primary variable (PV) out of limits, system failure in measurement circuit, deviation between A/D and D/A converter, check sum, watch dog and self-checking facility |
| Positions 0 ... 9, A ... F |
| Conforming to HART Communication Foundation (HCF) |
Certificates and approvals | |
General Purpose | CE, CSA, FM, RCM, KCC, EAC |
Non-incendive/Non-sparking | - CSA/FM Class I, Div. 2, Groups A, B, C, D T4 ATEX II 3G 2D EEx nA [ib] IIC
- T6 ... T4 T100 °C
|
Dust Ignition Proof (Intrinsically Safe Probe Circuit) | |
| CSA/FM Class II and III, Div. 1, Groups E, F, G |
| ATEX II 1/2 GD EEx d [ia] T6 ... T1 T100 °C |
| INMETRO Ex d [ia Ga] IIC T6 … T1 Gb Ex tb [ia Da] IIIC T100 °C Db -40 °C ? Ta ? +70 °C IP65/IP68 |
| EAC Ex |
Explosion Proof (Intrinsically Safe Probe Circuit) | |
| FM Class I, Div. 1, Groups A, B, C, D T4 |
| ATEX II 1/2 GD EEx d [ia] IIC T6 ... T1 |
Overfill Protection | AIB-Vincotte |
Marine | Lloyds Register of Shipping, Categories ENV1, ENV2, ENV3, and ENV5, Bureau Veritas |
Other | Pattern approval (AQSIQ, China), CRN |