C79165-A3044-B176
Внимание! Указаны цены архивного каталога СА01. Для уточнения актуальной цены перейдите по ссылке заказного номера товара или перейдите в каталог СА01 2016
Заказной номер | Вес | PG | Описание продукта | Цена |
C79165-A3044-B176 | 0.1 | 822 | TEMPERATURE SENSOR | 58.3 |
ULTRAMAT 6, 19" rack unit |
|
General information |
|
Measuring ranges |
4, internally and externally switchable; autoranging is also possible |
Smallest possible measuring range |
Dependent on the application: e.g. CO: 0 … 10 vpm, CO2: 0 … 5 vpm |
Largest possible measuring span |
Dependent on the application |
Measuring range with suppressed zero point |
Any zero point within 0 … 100 vol.% can be implemented; smallest possible span 20% |
Operating position |
Front wall, vertical |
Conformity |
CE mark in accordance with EN 50081-1, EN 50082-2 |
Influence of interfering gases must be considered separately |
|
Design, enclosure |
|
Weight |
Approx. 15 kg (with one IR channel) Approx. 21 kg (with two IR channels) |
Degree of protection |
IP20 according to EN 60529 |
Electrical characteristics |
|
EMC (ElectromagneticCompatibility) |
In accordance with standard requirements of NAMUR NE21 (08/98) |
Electrical safety |
According to EN 61010-1, overvoltage category III |
Power supply |
100 … 120 V AC (nominal range of use 90 … 132 V), 47 … 63 Hz or 200 … 240 V AC (nominal range of use 180 … 264 V), 47 … 63 Hz |
Power consumption |
1-channel unit: Approx. 40 VA 2-channel unit: Approx. 70 VA |
Fuse values |
|
|
1 T/250 (7MB2121) |
|
0.63 T/250 (7MB2121) |
Gas inlet conditions |
|
Permissible sample gas pressure |
|
|
|
|
600 … 1 500 hPa (absolute) |
|
700 … 1 300 hPa (absolute) |
|
600 … 1 500 hPa (absolute) |
Sample gas flow |
18 … 90 l/h (0.3 … 1.5 l/min) |
Sample gas temperature |
Min. 0 ... max. 50 °C, but above the dew point |
Sample gas humidity |
< 90% RH (relative humidity), or dependent on measuring task, non-condensing |
Dynamic response |
|
Warm-up period |
At room temperature < 30 min (the technical specification will be met after 2 hours) |
Delayed display (T90-time) |
Dependent on length of analyzer chamber, sample gas line and parameterizable damping |
Damping (electrical time constant) |
0 … 100 s, parameterizable |
Dead time (purging time of the gas path in the unit at 1 l/min) |
Approximately 0.5 ... 5 s, depending on version |
Time for device-internal signal processing |
< 1 s |
Pressure correction range |
|
Pressure sensor |
|
|
700 … 1 200 hPa absolute |
|
700 … 1 500 hPa absolute |
Measuring response(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
|
Output signal fluctuation |
< ± 1% of the smallest possible measuring range according to rating plate |
Zero point drift |
<± 1% of the current measuring range/week |
Measured-value drift |
<± 1% of the current measuring range/week |
Repeatability |
1% of the current measuring range |
Detection limit |
1 % of the smallest possible measuring range |
Linearity error |
< 0.5% of the full-scale value |
Influencing variables(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
|
Ambient temperature |
< 1% of current measuring range/10 K (with constant receiver cell temperature) |
Sample gas pressure |
|
Sample gas flow |
Negligible |
Power supply |
< 0.1% of the current measuring range with rated voltage ± 10% |
Environmental conditions |
Application-specific measuring influences possible if ambient air contains measured components or cross interference-sensitive gases |
Electrical inputs and outputs |
|
Analog output |
0/2/4 ... 20 mA, isolated; load 750 |
Relay outputs |
6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated, non-sparking |
Analog inputs |
2, dimensioned for 0/2/4 ... 20 mA for external pressure sensor and accompanying gas influence correction (correction of cross-interference) |
Binary inputs |
6, designed for 24 V, isolated, freely parameterizable, e.g. for measuring range switchover |
Serial interface |
RS 485 |
Options |
AUTOCAL function with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP |
Climatic conditions |
|
Permissible ambient temperature |
-30 ... +70 °C during storage and transportation, |
Permissible humidity |
< 90% RH (relative humidity) as annual average, during storage and transportation (dew point must not be undershot) |