Siemens
СРЕДСТВА ПРОМЫШЛЕННОЙ АВТОМАТИЗАЦИИ
официальный партнер Сименс
Каталог СА01 2012
архивный
(4872) 700-366
skenergo@mail.ru

A5E00852382


Внимание! Указаны цены архивного каталога СА01. Для уточнения актуальной цены перейдите по ссылке заказного номера товара или перейдите в каталог СА01 2016

Заказной номерВесPGОписание продуктаЦена
A5E008523821822USB-RS485 CONVERTER492.9

19" rack unit and portable version

ULTRAMAT 23, 19" rack unit and portable version

General information

Measured components

Maximum of 4, comprising three infrared-sensitive gases and oxygen

Measuring ranges

Two per measured component

Display

LCD with LED backlighting and contrast control; function keys; 80 characters (4 lines/20 characters)

Operating position

Front wall, vertical

Conformity

CE symbol EN 61000-6-2, EN 61000-6-4

Design, enclosure

Weight

Approximately 10 kg

Degree of protection, 19" rack unit and desktop model

IP20 according to EN 60529

Electrical characteristics

EMC (Electromagnetic Compatibility) (safety extra-low voltage (SELV) with safety isolation)

In accordance with standard requirements of NAMUR NE21 (08/98) or EN 50081-1, EN 50082-2

Power supply

100 V AC, +10%/-15%, 50 Hz,

120 V AC, +10%/-15%, 50 Hz,

200 V AC, +10%/-15%, 50 Hz,

230 V AC, +10%/-15%, 50 Hz,

100 V AC, +10%/-15%, 60 Hz,

120 V AC, +10%/-15%, 60 Hz,

230 V AC, +10%/-15%, 60 Hz

Power consumption

Approx. 60 VA

Electrical inputs and outputs

Analog output

Per component, 0/2/4 up to 20 mA, NAMUR, isolated, max. load 750

Relay outputs

8, with changeover contacts, freely parameterizable, e.g. for measuring range identification;

24 V AC/DC/1 A load, potential-free, non-sparking

Binary inputs

3, dimensioned for 24 V, potential-free

  • Pump

  • AUTOCAL

  • Synchronization

Serial interface

RS 485

AUTOCAL function

Automatic unit calibration with ambient air (depending on measured component); adjustable cycle time from 0 (1) …  24 hours

Options

Add-on electronics, each with 8 additional binary inputs and relay outputs for e.g. triggering of automatic calibration and for PROFIBUS PA or PROFIBUS DP

Climatic conditions

Permissible ambient temperature

  • During operation

5 … 45 °C

  • During storage and transportation

-20 ... +60 °C

Permissible ambient humidity

< 90% RH (relative humidity) during storage and transportation

Permissible pressure fluctuations

600 … 1 200 hPa

Gas inlet conditions

Sample gas pressure

  • Without pump

Unpressurized (< 1 200 hPa, absolute)

  • With pump

Depressurized suction mode, set in factory with 2 m hose at sample gas outlet; full-scale value calibration necessary under different venting conditions

Sample gas flow

72 … 120 l/h (1.2 … 2 l/min)

Sample gas temperature

Min. 0 to max. 50 °C, but above the dew point

Sample gas humidity

< 90% RH (relative humidity), non-condensing


Technical data, infrared channel

So that the technical data can be complied with, a cycle time of  24 hours must be activated for the AUTOCAL. The cycle time of the AUTOCAL function must be  6 hours when measuring small NO and SO2 measuring ranges (  400 mg/m ) on TUV/QAL-certified systems.

Measuring ranges

See ordering data

Chopper compartment flushing

Upstream pressure approximately 3 000 hPa; purging gas consumption approximately 100 ml/min

Dynamic response

Warm-up period

Approximately 30 min (at room temperature) (the technical specification will be met after 2 hours)

Delayed display (T90-time)

Dependent on length of analyzer chamber, sample gas line and parameterizable damping

Damping (electrical time constant)

Parameterizable from 0 … 99.9 s

Measuring response

(relating to sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature)

Output signal fluctuation

< ± 1% of the current measuring range (see rating plate)

Detection limit

1 % of the current measuring range

Linearity error

  • In largest possible measuring range:
    < ± 1% of the full-scale value
  • In smallest possible measuring range:
    < ± 2% of the full-scale value

Repeatability

± 1% of the current measuring range

Drift

Zero point

  • With AUTOCAL

Negligible

  • Without AUTOCAL

< 2 % of the current measuring range/week

Full-scale value drift

  • With AUTOCAL

Negligible

  • Without AUTOCAL

< 2 % of the current measuring range/week

Influencing variables

(relating to sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature)

Temperature

Max. 2% of the smallest possible measuring range according to rating plate per 10 K with an AUTOCAL cycle time of 6 h

Atmospheric pressure

< 0.2% of the current measuring range per 1% pressure change

Power supply

< 0.1% of the current measuring range with a change of ± 10%


Technical data, oxygen channel (electrochemical)

Measuring ranges

0 … 5 % … 0 … 25 % O2, parameterizable

Service life

Approximately 2 years at 21% O2; continuous duty < 0.5% O2 will destroy the measuring cell

Detection limit

1% of the current measuring range

Dynamic response

Delayed display (T90-time)

Dependent on dead time and parameterizable damping, not > 30 s at approximately 1.2 l/min sample gas flow

Measuring response

(relating to sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature)

Output signal fluctuation

< ± 0.5% of the current measuring range

Linearity error

< ± 0.2% of the current measuring range

Repeatability

0.05 % O2

Drift

  • With AUTOCAL

Negligible

  • Without AUTOCAL

1 % O2/year in air, typical

Temperature

< ± 0.5% O2 per 20 K, relating to a measured value at 20 °C

Atmospheric pressure

< 0.2 % of the measured value per 1 % pressure change

Influencing variables

(relating to sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature)

Oxygen content

Intermittent operation < 0.5% O2 leads to falsification of the measured value

Accompanying gases

The oxygen sensor must not be used if the accompanying gas contains the following components: Chlorine or fluorine compounds, heavy metals, aerosols, mercaptans, alkaline components (such as NH3 in % range)

Typical combustion exhaust gases

Influence: < 0.05% O2

Humidity

H2O dew point 2 °C; the oxygen sensor must not be used with dry sample gases (however, no condensation either)


Technical data, H2S channel

Measured components

Maximum of four, comprising one or two infrared-sensitive gases, one oxygen component and one hydrogen sulfide component

Measuring ranges of H2S sensor MB 5000

  • Smallest measuring range

0 ... 500 vpm

  • Largest measuring range

0 ... 5 000 vpm

Service life of the sensor

Approx. 12 months

Permissible ambient pressure

750 ... 1 200 hPa

Permissible operating temperature

5 ... 40 °C (41 ... 104 °F)

Influencing variables

Accompanying gases

The hydrogen sulfide sensor must not be used if the accompanying gas contains the following components:

  • Compounds containing chlorine
  • Compounds containing fluorine
  • Heavy metals
  • Aerosols
  • Alkaline components
    (e.g. NH3 > 5%)

Cross-inference (interfering gases)

100 ppm SO2 result in a cross-interference of < 30 ppm H2S

Drift

< 1% per month

Temperature

< 3%/10 K referred to full-scale value

Atmospheric pressure

< 0.2% of the measured value per 1% pressure change

Measuring response

Delayed display (T90 time)

< 80 s with sample gas flow of approx. 1 ... 1.2 l/min

Output signal noise

< 15 ppm H2S

Display resolution

< 0.2% of the full-scale value

Output signal resolution

< 30 ppm H2S

Repeatability

< 4% referred to full-scale value


Technical data, paramagnetic oxygen cell

Measured components

Maximum of four, comprising up to three infrared-sensitive gases and an oxygen component

Measuring ranges

Two per component

  • Min. 0 ... 2% vol O2
  • Max. 0 ... 100% vol O2

Permissible ambient pressure

700 ... 1 200 hPa

Permissible operating temperature

5 ... 45 °C (41 ... 113 °F)

Cross-inference (interfering gases)

See Table "Cross-sensitivities"

Zero point drift

  • Measuring range 2%: max. 0.1% with weekly zero adjustment
  • Measuring range 5%: max. 0.1% with weekly zero adjustment
  • Measuring range 25% or greater:
    max. 0.5% with monthly zero adjustment

Temperature error

< 2%/10 K referred to measuring range 5%

< 5%/10 K referred to measuring range 2%

Humidity error for N2 with 90% relative humidity after 30 min

< 0.6% at 50 °C

Atmospheric pressure

< 0.2% of measured value per 1% pressure change

Delayed display (T90 time)

< 60 s

Output signal noise

< 1% of smallest measuring range

Repeatability

< 1% of smallest measuring range
























































Каталог оборудования 2012
Каталог продуктов Siemens Industry Приводная техника Техника автоматизации Системы автоматизации Системы визуализации SIMATIC HMI Системы идентификации Промышленные коммуникации SIMATIC NET Промышленные аппараты управления SIRIUS Промышленные информационные технологии Управление на базе РС Системы управления процессом Датчики Контрольно-измерительные приборы Анализаторы процесса Непрерывные анализаторы газов Непрерывные анализаторы газов, эктрактивные Введение ULTRAMAT 23 General information 19" rack unit and portable version Documentation Suggestions for spare parts ULTRAMAT 6 ULTRAMAT/OXYMAT 6 OXYMAT 6 OXYMAT 61 OXYMAT 64 CALOMAT 6 CALOMAT 62 FIDAMAT 6 Непрерывный газовый анализ, in-situ Промышленные газовые хроматографы Analytical Application Sets General information Технологии взвешивания Датчики зрения Блоки питания Продукты для специальных требований Автоматизация зданий Низковольтная коммутационная техника Технология безопасности Системные решения и продукты для отраслей Сервис ... и все, что Вам еще необходимо

  © ООО "СК ЭНЕРГО" 2007-2022
  (4872) 700-366
Яндекс.Метрика